process yellow 116 1375 1797 375 3405 313 2985 240 Embedded Power for Business-Critical Continuity 242 2607 Rev.06.29.11 151 NLP250 Medical Series 1 of 4 652 NLP250 Medical Series Single Output Total Power: 250 W Input Voltage: 85 - 264 VAC of Outputs: Single Electrical Specifications Special Features Input Medical safeties Input voltage range: Universal input 85 - 264 Vac Active PFC and EN61000-3-2 compliant Input frequency range: 47 - 63 Hz 250 W on main channel with Input surge current: 264 Vac (cold start) 40 A max. forced air Safety ground leakage current: 264 Vac, 50 Hz 150 A Low profile fits 1U applications Input current: U-Channel for maximum 120 Vac 250 W 2.78 A rms thermal performance 230 Vac 250 W 1.36 A rms Optional cover (CJ suffix) T6.4 AH, 250 Vac In live and Input fuse: UL/IEC127 5 V standby output neutral 12 V fan output Output Integrated control and Maximum power: 200 LFM forced air 250 watts monitoring features 250 LFM with cover Overcurrent, overvoltage and Adjustment range: Main output 5% overtemperature protection Main output 2.0% Compliance to EN55022-B Total regulation: conducted noise standard (line and load) Auxiliary outputs 5.0% Dual AC fuses Turn-on delay: 120 Vac Input 2.0 s max. RoHS compliant Transient response: Main output 5.0% or 250 mV 2 year warranty 50 - 100% max. dev., 1 ms max Step at 0.5 A/s recovery to 1% Temperature coefficient: 0.02%/C Safety VDE0750/EN60950 Overvoltage protection: Main output 115%, 5% IEC950/IEC60601-1 Short circuit protection: Cyclic operation Continuous File No. 1177400-3336-0759 Minimum output current: Singles 0 A UL60601-1 Auxiliary outputs: 5 Vsb 5 V 1.0 A File No. E186249 (See Note 8, page 3) Certificate No. 40014041 12 V (fan) 12 V 0.3A CB Ref DE1-36628 All specifications are typical at nominal input, full load at 25 C unless otherwise statedprocess yellow 116 1375 1797 375 3405 313 2985 240 242 Embedded Power for Business-Critical Continuity 2607 Rev.06.29.11 151 NLP250 Medical Series 2 of 4 652 (5) EMC Characteristics Conducted emissions: EN55022, FCC part 15 Level B Harmonic current correction: EN61000-3-2 Compliant Voltage flicker: EN61000-3-3 Compliant ESD air: EN61000-4-2 Level 3 ESD contact: EN61000-4-2 Level 3 Radiated immunity: EN61000-4-3 Level 3 Fast transients: EN61000-4-4 Level 3 Surge: EN61000-4-5 Level 3 Conducted immunity: EN61000-4-6 Level 3 Voltage dips: EN61000-4-11 Compliant General Specifications Hold-up time: 85 Vac 50 Hz 20 ms 250 W Efficiency: 115 Vac 250 W 84% t y p. 230 Vac 250 W 86% t y p. Isolation voltage: Input/output 4000 Vac Input/chassis 2000 Vac Safety approvals (see Note 6, page 3): UL/cUL UL60601-1, VDE EN60601-1, CAN/CSA22.2 No. 601-1 Weight: 650g (22 oz) MTBF ( 25 C): Telcordia SR-332 317,000 hours min. M I L-HDB K-217F 158,000 hours min. Environmental Specifications Thermal performance: Operating ambient, 0 C to +70 C (See derating curve) Non-operating -40 C to +85 C 0 C to 50 C ambient, 250 W 200 LFM forced air 250 LFM with cover 0 C to 50 C ambient, 175 W 0 C to 40 C with cover Convection cooled 50 C to 70 C ambient, Derate linearly Convection cooled to 50% load Relative humidity: Non-condensing 5 - 95% RH Altitude: Operating 10,000 feet max. Non-operating 30,000 feet max. Vibration (See Note 7, page 3): 5 - 500 Hz 2.5 G rms peak Shock: Per MIL-STD-810E 516.4 Part IV