MB60S 60W Single Output Medical Series Features Ultra Small size of 2 x 3 x 1.0 For 1U Applications 60W convection cooled Universal Input 90-264Vac rd Approved to IEC60601-1, 3 Edition with 2 MOPP Level V Efficiency Compliant Models Less than 0.5W no-load Power Consumption 3 Year Warranty Optional LED indicator for power-on RoHS Compliant Description The MB60S Series models provide a reliable power source in high power density in 2 x 3 x 1.0 package. Fully compliant to the applicable safety and EMC standards, these models will allow easy integration into many Medical applications. All 6 models are CE marked to low voltage directive and approved to Medical standards rd of IEC60601-1 3 edition with 2 MOPP. Model Selection Model Output Current Output Power Ripple & Total OVP Number Volts Convection Cooled Convection Cooled Noise* Regulation Threshold MB60S12K 12V 4.58A 55W 120mV pk-pk 14.4-18Vdc 2% MB60S15K 15V 4.00A 60W 150mV pk-pk 18-22.5Vdc 2% MB60S18K 18V 3.33A 60W 180mV pk-pk 2% 21-25.5Vdc MB60S24K 24V 2.50A 60W 240mV pk-pk 2% 28.8-36Vdc MB60S36K** 36V 1.67A 60W 360mV pk-pk 2% 42-47Vdc MB60S48K 48V 1.25A 60W 480mV pk-pk 2% 57.6-72Vdc Notes: * At -20C, the noise and ripple is 2% of the output. ** For product availability, please contact the factory Input Specifications PARAMETER SPECIFICATION NOTES AC Input Voltage: 90-264Vac, single phase AC Input Frequency: 47-63Hz AC Input Current: 120Vac: 1.4A, 240Vac: 0.75A Turn-on Input Voltage: 70V Ramping Up Turn-off Input Voltage: 65V Ramping Down Inrush Current: 40A maximum 0C www.slpower.com MB60S 26-Aug-13 1 MB60S 60W Single Output Medical Series Leakage Current <275A 264Vac, 60 Hz input, NC rd (InputEarth): IEC 60601-1 3 Ed 8.7.3.c Leakage Current N/A (OutputEarth): Leakage Current <90A 264Vac, 60 Hz input, NC (Input-Output): Input Fuses: F1, F2: 4A, 250VAC Fuses provided on all models Efficiency Typical Measured at 120Vac and full load MB60S12K 83% MB60S15K 85% MB60S18K 85% MB60S24K 88% MB60S36K 88% 24V, 36V, and 48V Models meet Level V requirement MB60S48K 88% No Load Input Power: <0.5W Meet Level V, standby Power Consumption Turn-on Time: <2 Seconds at 120Vac. 16mS minimum from loss of ac input at Hold-up Time: 55 Watts for 12V output 120 Vac, full load. DC Output Specifications PARAMETER SPECIFICATION NOTES 60W continuous for operation from -10C Output Power: to 50C55 Watts for 12V output. Cooling: Convection Total regulation is the maximum deviation from Total Regulation: 2% for all models nominal voltage for all loading conditions 120% - 180% of rated output current Overload Protection: value, Hiccup Mode Short Circuit Short across the output terminals will not Protection: cause damage to the unit. Hiccup Mode OVP firing reduces output voltage to Overvoltage <50% of nominal in <50mS. See chart for Protection: trip range Overtemperature Automatic Power Shutdown at Tc = Protection: 155C, Minimum Load: No minimum load is required 20 MHz Bandwidth, differential mode. 0.5% RMS, 1% pk-pk for all models. Measured with noise probe directly across Ripple and Noise: output terminals, and load terminated with 0.1F ceramic and 10F low ESR capacitors 500s typ. response time for return to within 0.5% of final value for a 50% load Transient Response: change, i/t< 0.2A/s. Max. voltage deviation is 3.5%. 5% overshoot at turn-on, 5% overshoot at Overshoot: turn-off, under all conditions. www.slpower.com MB60S 26-Aug-13 2