MB65S 65W Single Output Medical Series Features 2 x 3.5 x 1.3 Package, Ideal for 1U Applications 10-year life design with Premium E-Caps 65 Watts convection cooled Class I and Class II models Class B Conducted and Radiated EMI Performance BF Isolation Type Rated th Designed to meet new IEC 60601-1-2 4 Edition EMC Requirements Approved to AAMI ES/CSA C22.2/EN/IEC60601-1, rd 3 Edition 2 x MOPP Isolation Internal Temperature Monitor, DC OK Signal, LED 3 Year Warranty Description A superior performance 65 Watt AC to DC power supply designed for next generation medical applications. Feature rich and highly efficient, MB65 product family can easily fit in a 1U chassis and provides 65 Watts of convection power. Input & output and internal temperature monitoring/alarms are features of the MB65 family. All models are CE marked to low rd voltage directive and approved to AAMI ES/CSA C22.2 No./EN/IEC60601-1, 3 edition. The design takes into th consideration the pending international release of the new IEC 60601-1-2, 4 Edition EMC requirements. With low leakage current performance, the power supplies are BF rated. Model Selection Model Output Current Total OVP 1 2 Number Class Volts (Convection Cooled) Efficiency Ripple &Noise Regulation Threshold MB65S12K Class I 12V 5.4A 88% 0.5% 1% 17.3 2.4V MB65S12C Class II 12V 5.4A 88% 0.5% 1% 17.3 2.4V MB65S15K Class I 15V 4.3A 88% 0.5% 1% 20.1 2.4V MB65S15C Class II 15V 4.3A 88% 0.5% 1% 20.1 2.4V MB65S24K Class I 24V 2.7A 89% 0.5% 1% 29.3 3.5V MB65S24C Class II 24V 2.7A 89% 0.5% 1% 29.3 3.5V MB65S48K Class I 48V 1.35A 90% 0.5% 1% 55.5 4.5V MB65S48C Class II 48V 1.35A 90% 0.5% 1% 55.5 4.5V Notes: 1) Efficiency, Typical at 115/230Vac, 25C. See Charts below for load conditions. 2) Measured at 25C using 6 inch twisted pair wires with noise probe directly across output terminals, and load terminated with 0.1F ceramic and 10F low ESR capacitors. 3) Part number suffix to include K for Class-1 AC input, or C suffix for Class-2 AC Input Consult Factory for following Models, availability may vary. Class I MB65S18K 18V 3.6A 89% 0.5% 1% 21.7 2.5V Class II MB65S18C 18V 3.6A 89% 0.5% 1% 21.7 2.5V Class I MB65S28K 28V 2.3A 89% 0.5% 1% 35.2 3.5V Class II MB65S28C 28V 2.3A 89% 0.5% 1% 35.2 3.5V Class I MB65S32K 32V 2.0A 89% 0.5% 1% 41.5 4.5V Class II MB65S32C 32V 2.0A 89% 0.5% 1% 41.5 4.5V www.slpower.com MB65S Aug -14 1 MB65S 65W Single Output Medical Series General Specifications 85-264Vac, single phase. AC Input Turn On Time <2 Seconds at 110Vac. (Safety Approved to 90-264Vac). Input Current 1.5A at 110Vac, 1A at 240Vac Rise Time <30mS, 48V model: < 40mS 40Arms Maximum within a half line cycle, cold 20mS min. from loss of AC input at 110 Vac, full load, Inrush Current Hold-up Time start at 25C. See application note. 25C. Over Load 115% - 200% of rated output current value. Hiccup IT Characteristic See Table below. Mode, Auto-recovery Protection (OCP) Turn-On Input Short Circuit Short across the output terminals will not cause >75Vac 8Vac. Full spec performance at 85Vac. damage to the unit. Hiccup Mode , Auto-recovery Voltage Protection (SCP) Turn-Off Input Over Voltage Latches off when output voltage is with range as <63Vac 8Vac shown in table. Requires AC Power cycle to reset Voltage Protection (OVP) Over Temperature Power shuts down at temperature of 70 (typical) Input Fuses 3.15A, 250Vac, line and neutral inputs at full load. Hiccup Mode , Auto-recovery Protection (OTP) Outputs protected against momentary reverse Earth Leakage Output Reverse current less than 20A peak for less than 10mS with <350 A 264Vac, 60Hz input, NC 0.5A average. Sustained reverse current at high Current Voltage Protection levels may damage unit. Patient Leakage Input-Output: 4500Vac, 2 x MOPP <4 mA 264Vac, 60Hz input, NC for BF rating Current (Output to Isolation Input-Ground: 1900Vac, 1 x MOPP Output-Ground: 1900Vac, 1 x MOPP Earth) <90 A 264Vac, 60 Hz input, NC, also suitable for -20C to +70C. Turn-On & Touch BF rating Turn on Temperature = -40C at >=120Vac, allowing Operating <450 A 264Vac, 60Hz input, SFC, also suitable x seconds for stabilization. Current Temperature for BF rating De-rated output power at 70C = 45.5W Storage <0.5W -40C to +85C No Load Input Power Temperature 88% - 90% typical at 115/230Vac, 25C. See chart Operating: -500m to 3000m Efficiency Altitude for additional details Non-operating: -500 to 40,000 feet 65W continuous convection cooled, -20C to 50C Output Power ambient. 85Vac to 264Vac. See chart for de- Relative Humidity 5% to 95%, non-condensing rating above 50C. Operating: Half-sine shock waveform. Impact Acceleration: 20g, Pulse duration: 11mS. 500S typ. response time for return to within Shock Cycles: 3 times per axis in X,Y, Z direction Transient Response 0.5% of final value for a 50% load change, i/t< Non-Operating: Half-sine shock waveform. Impact (IEC 60068-2-27) 0.2A/s. Max. volt. deviation is 3.5%. Acceleration: 100g, Pulse duration: 6mS Cycles: 3 times per direction on 3 axes (X,Y. Z) Operating: Sinusoidal Frequency: 10-500Hz, Impact Acceleration: 1g, Sweep rate: 1 octave/min Cycles: 10 times per axis in X, Y, Z direction Vibration Random Vibration: Ripple and Noise 0.5% pk-pk (IEC 60068-2-6) 2 Operating: 0.003g /Hz, 1.224grms overall, 3 axes, (IEC 60068-2-64) 10 min per axis, 1-500Hz. 2 Non-Operating: 0.02g /Hz, 3.1grms overall, 3 axes, 1 hour per axis, 20-500 Hz 12V to 48Vdc. See models chart for part 564,500 hours (ave) 110Vac, 25C, SR332 Issue 6. Output Voltage MTBF numbering. Consult Factory 220Vac, 25C, SR332 Issue 6. >10 Years in use condition of 40C ambient, at +/- 10%, 48V is +/- 5% E-Cap Life 12h/day, 261 days/year. Additional information on Voltage Adjustability other use profiles available on request. www.slpower.com MB65S Aug -14 2