MINT1150 150W Single Output Medical Series Features 2 x 4 x 1.3 Package For 1U Applications 150W w/air, 100W convection cooled Universal Input 90-264 VAC Power Fail/Output Good Signal Approved to CSA/EN/IEC/UL60601-1, rd 3 Edition 2 x MOPP Input to Output Isolation 3 Year Warranty RoHS Compliant Description The MINT1150 family is a high power density for a power supply in a 2 x 4 size. Approved to EN/IEC/UL60601-1, 3rd edition, with isolation levels which satisfy the 2 MOPP requirements. The MINT1150 family is ideal for Portable Medical Devices, and many other applications where medical certifications, power density and cost are critical. The MINT1150 operates at universal input range of 90 to 264Vac and wide temperature range -10C to +70C, delivering full rated output power up to +50C. In addition, these models feature Power Fail and DC OK signals. Model Selection Model Output Current Ripple & Total OVP Number Volts w/200LFM air Convection* Noise** Regulation Threshold** MINT1150A1206K01 12V 12.5A 8.33A 0.5%RMS, 1.2% pk-pk 5% 14.0 1.1V MINT1150A1506K01 15V 10.0A 6.67A 0.5%RMS, 1.0% pk-pk 5% 18.0 1.5V MINT1150A2406K01 24V 6.25A 4.17A 0.5%RMS, 1.0% pk-pk 5% 28.0 2.5V MINT1150A4806K01 48V 3.13A 2.08A 0.5%RMS, 1.0% pk-pk 5% 55.0 4.0V MINT1150A5606K01 56V 2.68A 1.79A 0.5%RMS, 1.0% pk-pk 5% <59.9V Notes: * Maximum output power is 95 Watts for input voltage of 90-105VAC at 50C convection. For input voltage of 105Vac or more, the total power is 100 Watts at 50C convection. ** Measured with noise probe directly across output terminals, and load terminated with 0.1F ceramic and 10F low ESR capacitors. General Specifications Less than 2 sec. 115Vac (inversely 100-240Vac, 10%, 47-63Hz, 1 AC Input Turn On Time proportional to input voltage and thermistor 120370Vdc temperature) Input Current 115Vac: 2A, 230Vac: 1A Hold-up Time >12mS at full load, 120Vac 264Vac, cold start: will not exceed 50A AC Power Fail, DC OK Inrush Current Signals www.slpower.com MINT1150 REV 1.4, 13-NOV-12 1MINT1150 150W Single Output Medical Series General Specifications (continued) Overload F1, F2: 4A, 250Vac fuses provided on all Input Fuses Hiccup Mode models Protection Earth Leakage Short Circuit Provided - no damage will occur if the output is <300A 264Vac, 60Hz, NC shorted. Hiccup Mode Current Protection Overvoltage OVP firing reduces output voltage to <50% of Efficiency 89% typical at 115Vac nominal in <50mS. See chart for trip range. Protection 150W continuous with 200 lfm airflow, 100W PFC: Variable 30-400kHz. Switching Output Power convection cooled See chart for specific Main Converter: Variable 35-180kHz, 65-70kHz Frequency voltage model ratings. at full load. Input-Output: 4000Vac, 2 x MOPP Transient 50% load step. i/ t: <0.2A/S Isolation Input-Ground: 1800Vac, 1 x MOPP Response Max Volt Deviation = 3% Output-Ground: 500Vac Operating See chart -10C to +70C Ripple and Noise Temperature Temperature Derate output power linearly above 50C to 50% Output Voltage See chart Derating at 70C Voltage Storage +/-5% from nominal -40C to +85C Adjustability Temperature Operating: -500 to 10,000 ft Not required Minimum Load Altitude Non-operating: -500 to 40,000 ft. Total Regulation +/- 5%. See chart Relative Humidity 5% to 95%, non-condensing Operating: 0.003g/Hz, 1.5grms overall, 3 Operating: Half-sine, 20gpk, 10ms, 3 axes, 6 axes, 10 min/axis shocks total Vibration Shock Non-Operating: 0.026g2/Hz, 5.0grms overall, Non-Operating: Half-sine, 40 gpk, 10 ms, 3 3 axes, 1 hr/axis axes, 6 shocks total rd Dimensions W: 2.0 x L: 4.0 x H: 1.3 Safety Standards EN/CSA/UL/IEC60601-1, 3 Edition 640,000 hours at 100W convection, 1,500,000 183g Weight MTBF hours at 150W with 200LFM air Auxiliary Signals Goes LOW with 5 mS warning before loss AC Power Fail: During normal operations, stays HIGH. Power Fail: of output power due to AC failure Open collector logic signal goes and stays HIGH 100mS to 500mS after main output reaches DC OK: regulation. www.slpower.com MINT1150 REV 1.4, 13-NOV-12 2