3
Very High Power Density: 15.27W/in
Class I or Class II Configuration
UL/EN60601-1 Medical Approval
Small 2x3 package
Efficiencies up to 91%
Suitable for BF Applied Part Applications
Meets Efficiency Level VI Requirements
No load power consumption <300mW
Electrical Specifications
Input Environmental
3
1 Operating Temperature -30C to +70C
Input Voltage 90-264VAC
Storage Temperature -30C to +85C
Input Frequency 47-63 Hz
Humidity 20% - 90% RH
Input Current (RMS) 2A max. @ 115VAC
Operating altitude <5,000m
1A max. @ 230VAC
MTBF: >250K hours per MIL-HDBK-217F at full load
Inrush Current 45A max @ 115VAC
and 25C ambient
90A max @ 230VAC
No Load Power Consumption <300mW
Power Factor >0.9 @ 240VDC and 120VDC
Safety & EMC
Safety
Output
Approvals
Total Output Power 120W with 10CFM Forced Air USA/Canada UL60601-1 3.1 Edition,
2
100W Convection Only UL62368-1 (Pending)
rd
Output Voltage See models and ratings table. Europe IEC/EN 60601-1 3 Edition, CB Report,
Hold Up Time 10mS minimum EN62368-1 (Pending)
Efficiency Up to 91%. See models and ratings table.
Line Regulation 1%
Isolation
5
Voltage Adjustability 4% 4000VAC / 5656VDC (2 x MOPP)
Input to Output
5
Setpoint accuracy 2% Input to Ground 2000VAC / 2828VDC (1 x MOPP)
5
Minimum Load No Minimum Load Output to Ground 1500VAC / 2121VDC (1 x MOPP)
<100A at 264Vac
Touch Current
Protection Features
EMC (IEC60601-1-2:2014):
EN55011 Class B
Emissions
Overvoltage Latch off.
EN55011 Class B (Class I)
Conducted
Overtemperature Latch off.
EN55011 Class A (Class II)
Radiated
Overcurrent Hiccup Mode. OCP Threshold typically
IEC/EN60601-1-2: 2007
Susceptibility
150%.
IEC 61000-3-2: Class A
Harmonic Currents
Short Circuit Hiccup Mode.
IEC 61000-3-3
Voltage Flicker
IEC 61000-4-2: 15kV Air, 8kV contact
Electrostatic Discharge
IEC 61000-4-3: 10V/m
Radiated Immunity
Notes
IEC 61000-4-4: +/-2kV
EFT/Burst
IEC 61000-4-5: 2005 1kV diff, 2kV com
Surge Immunity
1. Derate output power by 0.8%/V below 115VAC
IEC 61000-4-6: 10Vrms
Conducted Immunity
2. Derate convection only output power by 10% for U-Channel models.
IEC 61000-4-8: 30A/m
Magnetic Field
Derate convection only output power by 15% for Enclosed models.
IEC 61000-4-11: 30% reduction for 500ms,
Dips / Interruptions
3. See derating curves on the second page for operation above 50C
100% reduction for 10ms.
4. Dimensions given are those of PCB. I/O conductors extend slightly
Physical
beyond PCB edge.
5. Testing isolation with an AC generator is not recommended. Either
Dimensions 3 .04L x 2W x 1 .3 1 H Typical
perform isolation testing with a DC voltage, or consult with our
Weight 6.4oz Typical
engineering staff for AC test considerations.
Your Partners in Power ...
Power Partners, Inc. | 43 Broad Street | Hudson, MA 01749
Tel: (978)567-9600 | Fax: (978)567-9601
Specifications subject to change.
PDAM120: February 20, 2019
Website: www.powerpartners-inc.com
Medical
AC/DC Power Models and Ratings
6
Model Number Output Maximum Maximum Output Ripple & Max Typical
7
Voltage Load with Load with Load Noise (Vp-p) Capacitive Efficiency
Convection 10CFM Regulation Load at 230VAC
2
Cooling Forced Air (F)
PDAM120-12A 12V 8.333A 10.000A +/-1% 160mV 3000 90%
PDAM120-14A 24V 4.167A 5.000A +/-1% 240mV 1500 90%
PDAM120-18A 48V 2.083A 2.500A +/-1% 480mV 500 91%
PDAM120-12B 12V 7.500A 10.000A +/-1% 160mV 3000 90%
PDAM120-14B 24V 3.750A 5.000A +/-1% 240mV 1500 90%
PDAM120-18B 48V 1.875A 2.500A +/-1% 480mV 500 91%
PDAM120-12C 12V 7.083A 10.000A +/-1% 160mV 3000 90%
PDAM120-14C 24V 3.541A 5.000A +/-1% 240mV 1500 90%
PDAM120-18C 48V 1.770A 2.500A +/-1% 480mV 500 91%
Notes (continued from first page):
6. Model number ending with A indicates open frame format.
Model number ending with B indicates U-channel format.
Model number ending with C indicates enclosed format.
7. Measured at 20MHz bandwidth with a 47uF electrolytic and 0.1uF ceramic capacitor in parallel with the DC output rails.
8. Input voltage derating and thermal derating are superimposed. The PDAM120 cannot operate with input voltages below
99VAC in thermal environments below 10C
8
Derating Curves
Your Partners in Power ...
Power Partners, Inc. | 43 Broad Street | Hudson, MA 01749
Tel: (978)567-9600 | Fax: (978)567-9601
Specifications subject to change.
PDAM120: February 20, 2019
Website: www.powerpartners-inc.com
Medical